Table 5 The horizontal discrepancies according to implant position in the arch, using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
Table 5 The horizontal discrepancies according to implant position in the arch, using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
author: Motaz Osman, Hassan Ziada, Ahmed Suliman, Neamat Hassan Abubakr | publisher: drg. Andreas Tjandra, Sp. Perio, FISID
Horizontal discrepancies Position N Open tray Closed tray P value Mean Median S.D Mean Median S. D Maxilla Anterior 9 0.069 0.040 0.055 0.047 0.020 0.050 0.110 Posterior 9 0.098 0.040 0.093 0.101 0.050 0.090 0.136 Total 18 0.084 0.040 0.075 0.075 0.035 0.070 0.584 Mandible Anterior 4 0.030 0.035 0.022 0.063 0.079 0.030 0.999 Posterior 18 0.050 0.030 0.059 0.102 0.109 0.045 0.118 Total 22 0.0464 0.04 0.054 0.095 0.104 0.040 0.152
Table 5 The horizontal discrepancies according to implant position in the arch, using the Mann-Whitney U test
Serial posts:
-
Introduction : A prospective clinical study on implant impression accuracy
-
Materials and methods : A prospective clinical study on implant impression accuracy [1]
-
Materials and methods : A prospective clinical study on implant impression accuracy [2]
-
Materials and methods : A prospective clinical study on implant impression accuracy [3]
-
Results : A prospective clinical study on implant impression accuracy
-
Discussion : A prospective clinical study on implant impression accuracy [1]
-
Discussion : A prospective clinical study on implant impression accuracy [2]
-
Discussion : A prospective clinical study on implant impression accuracy [3]
-
Conclusion : A prospective clinical study on implant impression accuracy
-
Availability of data and materials : A prospective clinical study on implant impression accuracy
-
References : A prospective clinical study on implant impression accuracy [1]
-
References : A prospective clinical study on implant impression accuracy [2]
-
References : A prospective clinical study on implant impression accuracy [3]
-
Acknowledgements : A prospective clinical study on implant impression accuracy
-
Funding : A prospective clinical study on implant impression accuracy
-
Author information : A prospective clinical study on implant impression accuracy
-
Ethics declarations : A prospective clinical study on implant impression accuracy
-
Additional information : A prospective clinical study on implant impression accuracy
-
Rights and permissions : A prospective clinical study on implant impression accuracy
-
About this article : A prospective clinical study on implant impression accuracy
-
Table 1 The t test for horizontal measurements of the intraoral and master cast in the open and closed tray techniques : A prospective clinical study on implant impression accuracy
-
Table 2 Open and closed tray techniques accuracy using the Wilcoxon signed-rank test : A prospective clinical study on implant impression accuracy
-
Table 3 Open and closed tray technique accuracy in the maxilla and mandible, using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
-
Table 4 Impression technique accuracy in the anterior and posterior regions using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
-
Table 5 The horizontal discrepancies according to implant position in the arch, using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
-
Table 6 Chi-square test of marginal discrepancies for the impression techniques, by implant position in the arch : A prospective clinical study on implant impression accuracy
-
Fig. 1. Horizontal measurements between the two impression copings in the patient’s mouth : A prospective clinical study on implant
-
Fig. 2. Light cure acrylic resin verification jig in the patient’s mouth : A prospective clinical study on implant
-
Fig. 3. Sample distribution according to arch and position : A prospective clinical study on implant
-
Fig. 4. Normality line of the distribution horizontal measurement data for the intraoral and working casts : A prospective clinical study on implant
-
Fig. 5. Marginal discrepancy distribution in the open and closed techniques, maxillary mandibular, and anterior and posterior regions : A prospective clinical study on implant
| Horizontal discrepancies | Position | N | Open tray | Closed tray | P value | ||||
|---|---|---|---|---|---|---|---|---|---|
| Mean | Median | S.D | Mean | Median | S. D | ||||
| Maxilla | Anterior | 9 | 0.069 | 0.040 | 0.055 | 0.047 | 0.020 | 0.050 | 0.110 |
| Posterior | 9 | 0.098 | 0.040 | 0.093 | 0.101 | 0.050 | 0.090 | 0.136 | |
| Total | 18 | 0.084 | 0.040 | 0.075 | 0.075 | 0.035 | 0.070 | 0.584 | |
| Mandible | Anterior | 4 | 0.030 | 0.035 | 0.022 | 0.063 | 0.079 | 0.030 | 0.999 |
| Posterior | 18 | 0.050 | 0.030 | 0.059 | 0.102 | 0.109 | 0.045 | 0.118 | |
| Total | 22 | 0.0464 | 0.04 | 0.054 | 0.095 | 0.104 | 0.040 | 0.152 | |
Table 5 The horizontal discrepancies according to implant position in the arch, using the Mann-Whitney U test
- Introduction : A prospective clinical study on implant impression accuracy
- Materials and methods : A prospective clinical study on implant impression accuracy [1]
- Materials and methods : A prospective clinical study on implant impression accuracy [2]
- Materials and methods : A prospective clinical study on implant impression accuracy [3]
- Results : A prospective clinical study on implant impression accuracy
- Discussion : A prospective clinical study on implant impression accuracy [1]
- Discussion : A prospective clinical study on implant impression accuracy [2]
- Discussion : A prospective clinical study on implant impression accuracy [3]
- Conclusion : A prospective clinical study on implant impression accuracy
- Availability of data and materials : A prospective clinical study on implant impression accuracy
- References : A prospective clinical study on implant impression accuracy [1]
- References : A prospective clinical study on implant impression accuracy [2]
- References : A prospective clinical study on implant impression accuracy [3]
- Acknowledgements : A prospective clinical study on implant impression accuracy
- Funding : A prospective clinical study on implant impression accuracy
- Author information : A prospective clinical study on implant impression accuracy
- Ethics declarations : A prospective clinical study on implant impression accuracy
- Additional information : A prospective clinical study on implant impression accuracy
- Rights and permissions : A prospective clinical study on implant impression accuracy
- About this article : A prospective clinical study on implant impression accuracy
- Table 1 The t test for horizontal measurements of the intraoral and master cast in the open and closed tray techniques : A prospective clinical study on implant impression accuracy
- Table 2 Open and closed tray techniques accuracy using the Wilcoxon signed-rank test : A prospective clinical study on implant impression accuracy
- Table 3 Open and closed tray technique accuracy in the maxilla and mandible, using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
- Table 4 Impression technique accuracy in the anterior and posterior regions using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
- Table 5 The horizontal discrepancies according to implant position in the arch, using the Mann-Whitney U test : A prospective clinical study on implant impression accuracy
- Table 6 Chi-square test of marginal discrepancies for the impression techniques, by implant position in the arch : A prospective clinical study on implant impression accuracy
- Fig. 1. Horizontal measurements between the two impression copings in the patient’s mouth : A prospective clinical study on implant
- Fig. 2. Light cure acrylic resin verification jig in the patient’s mouth : A prospective clinical study on implant
- Fig. 3. Sample distribution according to arch and position : A prospective clinical study on implant
- Fig. 4. Normality line of the distribution horizontal measurement data for the intraoral and working casts : A prospective clinical study on implant
- Fig. 5. Marginal discrepancy distribution in the open and closed techniques, maxillary mandibular, and anterior and posterior regions : A prospective clinical study on implant