Open hour: senin - sabtu 09:00:00 - 20:00:00; minggu & tanggal merah tutup

Abbreviations : Does the manual insertion torque o...

Resonance frequency analysis Implant stability quotient

Availability of data and materials : Does the manu...

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Conclusions : Does the manual insertion torque of ...

The null hypothesis that finger tightening of transducer devices for RFA analysis may have a negative impact on the accuracy of ISQ measurements was rejected. Different forces applied during transducer tightening by different clinicians had no significant effect on the resulting ISQ values. Manual tightening of smartpeg magnetic devices allows for an objective and reliable determination of ISQ val...

Discussion : Does the manual insertion torque of s...

A great number of studies have summarized that the measurement of implant stability with RFA is reliable, is noninvasive, and can be used at any time after implant insertion and during follow-up [19, 20]. Various parameters have been demonstrated to influence the degree of primary implant stability [21, 22]. These include bone density [23, 24], surgical technique [25], implant insertion torque [26...

Results : Does the manual insertion torque of smar...

In the test group, mesial and buccal ISQ values of 30 implants were recorded by four examiners (S1–S4) after hand tightening the smartpegs. The control group considered mesial and buccal ISQ values of 30 implants with mechanically inserted smartpegs. Only 147 values could be recorded because one transducer broke and thereafter no further values could be obtained for this implant. At torque value...

Material and methods : Does the manual insertion t...

Statistical analysis was performed to evaluate the difference between the test and control group. Variance analysis and continuous variables were determined. Pearson r correlation coefficient was tested to verify the relationship between the two variables (e.g., relationship between torque and ISQ). Statistical significance was set at p > 0.05.

Material and methods : Does the manual insertion t...

Three fresh bovine ribs from the same animal were selected for the current in vitro testing; the bovine ribs were of a similar size to those used by Gecikli et al. [16], thus attempting to imitate human edentulous bone with a similar composition of cortical and cancellous bone. The animal was farmed and sacrificed for food production. The bone was stored airtight, humid, and cool from the time the...

Background : Does the manual insertion torque of s...

While dental implants have become increasingly important in the functional and esthetic rehabilitation of patients, implant failure still does occur. Primary stability at the time of implant placement and the development of osseointegration in the following healing process (secondary stability) are essential parameters for implant success [1]. Primary stability can be considered as the biomechanic...

Abstract : Does the manual insertion torque of sma...

There is disagreement about the optimal torque for tightening smartpegs for resonance frequency analysis (RFA). Subjective finger pressure during hand tightening could affect the reliability of the resulting values. The aim of the current study was therefore to assess whether or not the insertion torque of a smartpeg magnetic device influences the implant stability quotient (ISQ) value during RFA....