Conclusions
A fractured abutment screw of a crown supported by an endosseous implant is an unhappy complication. The fragment retained in the fixture can be retrieved with a festooned #557 latch slow-speed burr set in reverse torque. Once the fragment is loosened and advanced coronally, a clockwise-turning #33 1/3 burr is touched to the fragment side to completely remove it. Care is taken n...
Discussion
A common complication of implant prosthetics is with components. The abutment screw connects the abutment to the fixture and can fracture under occlusal functional load.
In preparation for treatment of a patient presenting with an abutment screw fracture, a radiograph should be taken to ensure osseous support quality of the supporting implant. The fixture screw hole should be flush...
Introduction
Dental implants have been a life-enhancing modality for partially and completely edentulous patients. Implants can successfully support a cemented or screw-retained single crown. However, this modality is not without complications: A fractured abutment screw may occur when the prosthesis is under functional cyclic loading. The abutment screw may be overloaded and fracture, leaving th...