Table 2 One-way ANOVA for the difference in removal torque values among groups with different numbers of mechanical loading cycles (Of: Removal torque pattern of a combined cone and octalobule index implant)
Table 2 One-way ANOVA for the difference in removal torque values among groups with different numbers of mechanical loading cycles (Of: Removal torque pattern of a combined cone and octalobule index implant)
author: Kanyarin Benjaboonyazit, Pisaisit Chaijareenont, Pathawee Khongkhunthian | publisher: drg. Andreas Tjandra, Sp. Perio, FISID
Sum of squares
df
Mean square
F
p values
RTVs
Between groups
670.997
13
51.615
6.110
.000**
Within groups
591.291
70
8.447
Total
1262.287
83
Table 2 One-way ANOVA for the difference in removal torque values among groups with different numbers of mechanical loading cycles
Serial posts:
-
Abstract : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
-
Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
-
Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
-
Materials and methods : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
-
Materials and methods : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
-
Results : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
-
Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
-
Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
-
Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [4]
-
Conclusion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
Abbreviations : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
-
References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
-
References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
-
References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [4]
-
Acknowledgements : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
Author information : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
Ethics declarations : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
Rights and permissions : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
About this article : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
-
Table 1 Mean removal torque values (N cm) of the abutment screws of the implant-abutment connections in all groups (Of: Removal torque pattern of a combined cone and octalobule index implant)
-
Table 2 One-way ANOVA for the difference in removal torque values among groups with different numbers of mechanical loading cycles (Of: Removal torque pattern of a combined cone and octalobule index implant)
-
Table 3 Post hoc Tukey’s HSD test of mean removal torque values in all groups (Of: Removal torque pattern of a combined cone and octalobule index implant)
-
Fig. 1. Cone connection combined with octalobular index (Octatorx) : Removal torque pattern of a combined cone and octalobule index implant
-
Fig. 2. Schematic of test setup according to The ISO 14801 recommendations : Removal torque pattern of a combined cone and octalobule index implant
-
Fig. 3. Implant assembly embedded in resin block : Removal torque pattern of a combined cone and octalobule index implant
-
Fig. 4. Tightening the abutment screw and measuring the RTVs : Removal torque pattern of a combined cone and octalobule index implant
-
Fig. 5. Specimen mounted in a 30°-angled steel holder in ElectroPuls E1000 dynamic testing machine : Removal torque pattern of a combined cone and octalobule index implant
-
Fig. 6. Specimen mounted in a 30°-angled steel holder : Removal torque pattern of a combined cone and octalobule index implant
-
Fig. 7. Schematic diagram for experimental procedure : Removal torque pattern of a combined cone and octalobule index implant
-
Fig. 8. Change in RTVs according to the numbers of mechanical loading cycles : Removal torque pattern of a combined cone and octalobule index implant
| Sum of squares | df | Mean square | F | p values | ||
|---|---|---|---|---|---|---|
| RTVs | Between groups | 670.997 | 13 | 51.615 | 6.110 | .000** |
| Within groups | 591.291 | 70 | 8.447 | |||
| Total | 1262.287 | 83 | ||||
Table 2 One-way ANOVA for the difference in removal torque values among groups with different numbers of mechanical loading cycles
- Abstract : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
- Materials and methods : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- Materials and methods : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- Results : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [4]
- Conclusion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Abbreviations : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [4]
- Acknowledgements : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Author information : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Ethics declarations : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Rights and permissions : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- About this article : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Table 1 Mean removal torque values (N cm) of the abutment screws of the implant-abutment connections in all groups (Of: Removal torque pattern of a combined cone and octalobule index implant)
- Table 2 One-way ANOVA for the difference in removal torque values among groups with different numbers of mechanical loading cycles (Of: Removal torque pattern of a combined cone and octalobule index implant)
- Table 3 Post hoc Tukey’s HSD test of mean removal torque values in all groups (Of: Removal torque pattern of a combined cone and octalobule index implant)
- Fig. 1. Cone connection combined with octalobular index (Octatorx) : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 2. Schematic of test setup according to The ISO 14801 recommendations : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 3. Implant assembly embedded in resin block : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 4. Tightening the abutment screw and measuring the RTVs : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 5. Specimen mounted in a 30°-angled steel holder in ElectroPuls E1000 dynamic testing machine : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 6. Specimen mounted in a 30°-angled steel holder : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 7. Schematic diagram for experimental procedure : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 8. Change in RTVs according to the numbers of mechanical loading cycles : Removal torque pattern of a combined cone and octalobule index implant