Number of cycles | Subset for alpha = 0.05 | |||
---|---|---|---|---|
N | 1 | 2 | 3 | |
0 | 6 | −27.6667 | ||
50,000 | 6 | −21.7667 | ||
100,000 | 6 | −20.0333 | −20.0333 | |
800,000 | 6 | −19.9333 | −19.9333 | |
150,000 | 6 | −19.4500 | −19.4500 | |
1,800,000 | 6 | − 19.1583 | −19.1583 | |
600,000 | 6 | −19.0667 | −19.0667 | |
1,600,000 | 6 | −19.0417 | −19.0417 | |
1,400,000 | 6 | −18.2750 | −18.2750 | |
1,000,000 | 6 | −17.7750 | −17.7750 | |
400,000 | 6 | −17.3667 | −17.3667 | |
200,000 | 6 | −17.1000 | −17.1000 | |
1,200,000 | 6 | −15.9917 | −15.9917 | |
2,000,000 | 6 | −15.8417 | ||
p values | 1.000 | .055 | .424 |
Table 3 Post hoc Tukey’s HSD test of mean removal torque values in all groups
Serial posts:
- Abstract : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- Background : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
- Materials and methods : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- Materials and methods : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- Results : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
- Discussion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [4]
- Conclusion : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Abbreviations : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [1]
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [2]
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [3]
- References : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [4]
- Acknowledgements : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Author information : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Ethics declarations : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Rights and permissions : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- About this article : Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
- Table 1 Mean removal torque values (N cm) of the abutment screws of the implant-abutment connections in all groups (Of: Removal torque pattern of a combined cone and octalobule index implant)
- Table 2 One-way ANOVA for the difference in removal torque values among groups with different numbers of mechanical loading cycles (Of: Removal torque pattern of a combined cone and octalobule index implant)
- Table 3 Post hoc Tukey’s HSD test of mean removal torque values in all groups (Of: Removal torque pattern of a combined cone and octalobule index implant)
- Fig. 1. Cone connection combined with octalobular index (Octatorx) : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 2. Schematic of test setup according to The ISO 14801 recommendations : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 3. Implant assembly embedded in resin block : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 4. Tightening the abutment screw and measuring the RTVs : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 5. Specimen mounted in a 30°-angled steel holder in ElectroPuls E1000 dynamic testing machine : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 6. Specimen mounted in a 30°-angled steel holder : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 7. Schematic diagram for experimental procedure : Removal torque pattern of a combined cone and octalobule index implant
- Fig. 8. Change in RTVs according to the numbers of mechanical loading cycles : Removal torque pattern of a combined cone and octalobule index implant