Results : Customized SmartPeg for measurement of resonance frequency of mini dental implants

Results : Customized SmartPeg for measurement of resonance frequency of mini dental implants

author: Jagjit Singh Dhaliwal, Rubens F Albuquerque Jr, Ali Fakhry, Sukhbir Kaur, Jocelyne S Feine | publisher: drg. Andreas Tjandra, Sp. Perio, FISID

The ISQ values obtained while calibrating the customized SmartPeg were similar to in vivo results. Median ISQ values at insertion and at 6 postoperative weeks were 53.3 (IQR 8.3) and 60.5 (5.5) for the 3M™ESPE™MDIs, and 58.5 (4.75) and 65.5 (9.3) for the Ankylos® implants, respectively, with no statistical difference (Figs. 2 and 3). The ISQ values of both 3M™ESPE™ MDI and Ankylos® (Figs. 2 and 3) increased significantly from the time of insertion to 6-week post-insertion (p < 0.05).