Table 1 Comparison of standard and over-dimensioned protocol
Table 1 Comparison of standard and over-dimensioned protocol
author: Henning Staedt,Peer W Kmmerer,Elisabeth Goetze,Daniel G E Thiem,Bilal Al-Nawas, Diana Heimes | publisher: drg. Andreas Tjandra, Sp. Perio, FISID
Drilling sequence — standard protocol | Drilling sequence — over-dimensioned protocol | |||||
---|---|---|---|---|---|---|
n | Mean (SD) | CI | n | Mean (SD) | CI | |
ISQ | 45 | 68.33 (6.83) | 66.14–70.51 | 30 | 68.5 (8.82) | 65.08–71.92 |
IT (Ncm) | 45 | 90.56 (31.27) | 80.56–100.56 | 30 | 63.74 (48.61) | 44.89–82.59 |
TO (Ncm) | 45 | 93.59 (32.3) | 83.27–103.92 | 30 | 58.35 (40.43) | 42.67–74.02 |
ISQ implant stability quotient, IT insertion torque, TO torque out, SD standard deviation, CI confidence interval
Serial posts:
- Implant primary stability depending on protocol and insertion mode — an ex vivo study
- Background : Implant primary stability depending on protocol and insertion mode (1)
- Background : Implant primary stability depending on protocol and insertion mode (2)
- Background : Implant primary stability depending on protocol and insertion mode (3)
- Methods : Implant primary stability depending on protocol and insertion mode (1)
- Methods : Implant primary stability depending on protocol and insertion mode (2)
- Results : Implant primary stability depending on protocol and insertion mode
- Discussion : Implant primary stability depending on protocol and insertion mode (1)
- Discussion : Implant primary stability depending on protocol and insertion mode (2)
- Discussion : Implant primary stability depending on protocol and insertion mode (3)
- Table 1 Comparison of standard and over-dimensioned protocol
- Table 2 Comparison of manual and machine-driven insertion mode
- Table 4 Comparison of implant length
- Table 3 Comparison of implant diameter
- Figure 1. Implant types and drilling protocol
- Figure 2. Over-dimensioned protocol
- Figure 3. Comparison of standard and over-dimensioned protocol