Figure 5. The effect of grade 4 and grade 5 implan...

  Figure 5. The effect of grade 4 and grade 5 implant particles on human gingival fibroblast viability in vitro. Viability was determined using an MTT metabolic activity assay. Cells were exposed to either a dissolution products (ions and nanoparticles) or b culture medium containing suspended implant particles throughout the duration of the culture period. Cells were exposed to various con...

Figure 4. Titanium (Ti) and vanadium (V) content i...

  Figure 4. Titanium (Ti) and vanadium (V) content in Dulbecco’s Modified Eagle Medium (DMEM). a, b Dissolution products (media filtered through 0.2 μm PTFE membrane following initial soaking of the particles for 3 days) and c, d DMEM sampled during cell culture studies where cells were cultured with the particles over a period of 10 days (particles removed prior to ICP measurem...

Figure 3. Titanium (Ti) and vanadium (V) release f...

  Figure 3. Titanium (Ti) and vanadium (V) release from the particles in simulated body fluid (SBF). Experimental duration was 10 days. Results presented as mean ± standard deviation, n = 3

Figure 2. EDX spectra of particles produced by the...

  Figure 2. EDX spectra of particles produced by the mock implantoplasty procedure (SEM images in Fig. 1). a, b Particles from grade 4 commercially pure titanium implant, a angular microparticles and b small spheres. c, d particles from grade 5 titanium alloy, c angular microparticles and d small spheres (grade 5)    

Figure 1. Representative photo of implants and SEM...

  Figure 1. Representative photo of implants and SEM images of particles produced by mock implantoplasty procedure. a–c Straumann 021.4512, bone level, diameter 4.1 mm, regular CrossFit®, SLA® 12 mm Roxolid® (commercially pure grade 4 titanium). d–f Biohorizons PBR 50105, RBT 5.0 × 10.5 mm, 5.7 Platform (grade 5 titanium alloy). Arrows indicate titanium oxide spheres. Scale ba...

Discussion : Particle release from implantoplasty ...

Further, the effect of implant particles on other cell types within the oral and systemic environments should not be overlooked. Although the health hazards of FPs and NPs are relatively less well established, literature in the fields of toxicology does indicate a glimpse of possible toxicity that should compel clinicians to carefully weigh the possible adverse human health effec...

Discussion : Particle release from implantoplasty ...

This is due to the lack of vanadium and possibly larger particle size. Here, G5 particles are. Although a range of particle size of G5 was measured in this study, a portion of the particles generated from the mock implatoplasty process in the current study is comparable to that reported by Pioletti et al.. The internalisation of G5 particles, especially sub-micron particles, and the su...

Discussion : Particle release from implantoplasty ...

There was no distinct difference in the amount of titanium ions released from G4 and G5 particles. Direct exposure to G5 implant particles in culture did result in significantly reduced cell viability at all-time points, from 3 to 10 days of culture, while G4 implant particles demonstrated no adverse effect on cell viability (Fig. 5b). The cytotoxic effects of vanadium are well docum...

Discussion : Particle release from implantoplasty ...

Discussion Unalloyed titanium, often referred to as commercially pure grade 4 titanium (CpTi), usually contains some trace elements of carbon, oxygen, nitrogen and iron (American Society for Testing and Materials international standards). These trace elements improve the mechanical properties of CpTi and are found in higher amounts from grade 1 to 4 CpTi. Many dental implants are made from ...

Results : Particle release from implantoplasty of ...

Results Particles released from implants following the mock implantoplasty procedure were collected, and microparticle size of particles produced from the grade 4 (G4) and grade 5 (G5) implants was 77.4 ± 9.1 μm (modal number 66.3 μm) and 48.4 ± 6.4 μm respectively (modal number 43.1 μm). DLS analysis showed nano-sized particles were also present: hydrodynamic diameters were 125.4...

Materials & methods : Particle release from implan...

The media were filtered through 0.2 μm PTFE membrane syringe filters following incubation to remove the particles before use in cell culture. Group 2 (particle): Sterilised grade 4 and 5 Ti particles were suspended in DMEM at concentrations of 0.75, 1.5 and 3 mg ml−1 and used for cell culture without filtering. Basal DMEM and DMEM containing unprocessed grade 4 and 5 implants were used a...

Materials & methods : Particle release from implan...

Ion release from titanium particles Simulated body fluid (SBF) was chosen as the dissolution test solution as we were interested in what happens when the particles become embedded in the soft/hard tissue rather than their interaction with saliva. SBF was prepared using the Kokubo method. Seven hundred millilitres of deionised (DI) water in a 1-L polypropylene beaker was warmed to 37 °C in a wa...

Materials & methods : Particle release from implan...

Materials and methods Materials Reagents and solvents were purchased from Sigma-Aldrich (Dorset UK). Commercially pure grade 4 titanium implants (n = 3) were purchased from Straumann (Sussex UK, Model number 021.4512, bone-level implant diameter 4.1 mm, Regular CrossFit®, SLA® 12 mm Roxolid®) (Fig. 1a). Grade 5 Ti-6Al-4 V titanium alloy implants were purchased from Biohorizons (Berk...

Background : Particle release from implantoplasty ...

However, if previous records are available, then the diagnosis can be made with any increase in pocket depth with post remodelling bone loss of greater than 0.5 mm in the presence of bleeding and/or suppuration on gentle probing, A number of studies suggested this inflammatory disease is associated with anaerobic plaque bacteria. It has also been suggested peri-implantitis can also be related to...

Background : Particle release from implantoplasty ...

Background Dental implants offer a viable long-term treatment option for patients with missing teeth. The use of metallic dental implants has relatively high reliability and long-term success rates; however, it is not without complications and the need for ongoing maintenance persists. Particles are generated during the life span of an implant, and this can have significant physiological implic...

Particle release from implantoplasty of dental imp...

Abstract Background With increasing numbers of dental implants placed annually, complications such as peri-implantitis and the subsequent periprosthetic osteolysis are becoming a major concern. Implantoplasty, a commonly used treatment of peri-implantitis, aims to remove plaque from exposed implants and reduce future microbial adhesion and colonisation by mechanically modifying the implant sur...

Fibroblas

Fibroblas adalah suatu jenis sel yang bertanggung untuk memproduksi matriks ekstraseluler dan kolagen. Fibroblas memainkan peran penting dalam perbaikan jaringan. Fibroblas adalah sel jaringan ikat yang utama dalam tubuh manusia. Fibroblas adalah anggota family sel jaringan ikat bersama-sama dengan sel tulang rawan, sel tulang, sel lemak, dan sel otot polos. Sebagian kelas fibroblas dapat beruba...

FGF (Fibroblast growth factor)

FGF atau Fibroblast Growth Factor pertama ditemukan oleh Armelin dalam ekstrak pituitari pada tahun 1973. Family FGF punya massa molekuler antara 17 kDa hingga 34 kDa. Anggota family ini punya kesamaan identitas asam amino antara 13% sampai 71%. FGF sangat terkonservasi pada struktur gen dan sekuen asam amino. Selama perkembangan embrio, anggota-anggota family FGF punya peran yang berbeda-b...