Fig. 9. The comparison of two groups at average voxel values for each part. The comparison of voxel values by insertion torque. All specimens were classified into two groups by insertion torque
Fig. 8. The relationship between average voxel value and insertion torque (averaged over the entire treatment area). The comparison of average voxel value among IT groups. Average voxel value was 384.0 ± 154.6 in the low IT group, 387.7 ± 147.7 in the medium IT group, and 619.2 ± 200.4 in the high IT group
Fig. 8. The relationship between average voxel value and insertion torqu...
Fig. 7. The average voxel value between the maxilla and mandible. There was no difference between the maxilla (430.9 ± 211.6) and the mandible (475.6 ± 211.5) in the average voxel value. Also, no difference was found in each part
Fig. 7. The average voxel value between the maxilla and mandible. There was no difference between the maxilla (430.9 ± 211.6) and the mandible (475.6...
Fig. 6. The relationship between ISQ and insertion torque. Percentage of specimens showing ISQ ≥ 73 compared with groups by week. In all groups, a period of rapidly increasing percentages was observed (8–12 weeks in the low IT group, 4–6 weeks in the medium and high IT groups). In the medium and high IT Group, a statistically significant difference was observed between ISQ ≥ 73 a...
Fig. 5. The comparison of ISQ values by the insertion torque. Time-lapse migration of ISQ values was compared with IT groups. Each IT group displayed similar migration. A significant difference in The ISQ was found in the low IT group after 8 weeks
Fig. 5. The comparison of ISQ values by the insertion torque. Time-lapse migration of ISQ values was compared with IT groups. Each IT group displ...
Fig. 4. The classification of the insertion torque. All specimens classified into three groups according to insertion torque. Criteria for the classification are shown in the figure and in the “Methods” section
Fig. 4. The classification of the insertion torque. All specimens classified into three groups according to insertion torque. Criteria for the classification are shown in the figur...
Fig. 3. The evaluation of the average ISQ. Time-lapse migration of average ISQ. Average ISQ of all specimens increased in a time-dependent manner (results indicated by a line). A significant difference was observed by 6 weeks after surgery
Fig. 3. The evaluation of the average ISQ. Time-lapse migration of average ISQ. Average ISQ of all specimens increased in a time-dependent manner (results...
Fig. 2. The measurement of the voxel values. A case of bone quality diagnosis before treatment. Width and height of the bone were measured to select the proper size of the implant body. The selected implant body was simulated on the bone images as a symbol, and then the voxel value was calculated as described in the “Methods” section
Fig. 2. The measurement of the voxel values. A case of ...
Fig. 1. Genesio® Plus implant with Aanchor surface. Scheme of the dental implant body for the Genesio® Plus implants with Aanchor surface used. a Overview picture of Genesio® Plus implants with Aanchor surface. b Image from scanning electron microscopy. Both pictures were provided by GC Corporation. To obtain osseointegration from an early stage, the dental implant body was treated with sandb...
Number of implants
Insertion torque value (N cm)
Implant stability quotient value
0 week
2 weeks
...
Number of implants
Treatment area (FDI)
Size of implant (mm)
Length
Diameter
1
...
Homma, S., Makabe, Y., Sakai, T. et al. Prospective multicenter non-randomized controlled study on intraosseous stability and healing period for dental implants in the posterior region.
Int J Implant Dent 4, 10 (2018). https://doi.org/10.1186/s40729-018-0122-x
Download citation
Received: 05 September 2017
Accepted: 25 January 2018
Published: 29 March 2018
DOI: https://doi...
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were m...
This research was conducted in accordance with the Helsinki Declaration of 1964 and subsequent ethical standards, with the approval of Tokyo Dental University and Fukuoka Dental College Ethics Committee (World Medical 2013). The surgeon explained the contents of the research to each patient, who then consented to participate in the present study.
All participants consented to publish their inform...
Correspondence to
Shinya Homma.
Department of Oral and Maxillofacial Implantology, Tokyo Dental College, 2-9-18 Misaki-cho, Chiyoda-ku, Tokyo, 101-0061, Japan
Shinya Homma, Yasushi Makabe & Yasutomo Yajima
Section of Oral Implantology, Department of Oral Rehabilitation, Fukuoka Dental College, 2-15-1 Tamura, Sawara-ku, Fukuoka-City, Fukuoka, 814-0175, Japan
Takuya Sakai, Kenzou Morinaga & Hirofumi Kido
Center for Oral ...
Nedir R, Bischof M, Szmukler-Moncler S, Bernard JP, Samson J. Predicting osseointegration by means of implant primary stability. Clin Oral Implants Res. 2004;15(5):520–8.
Ito Y, Sato D, Yoneda S, Ito D, Kondo H, Kasugai S. Relevance of resonance frequency analysis to evaluate dental implant stability: simulation and histomorphometrical animal experiments. Clin Oral Implants Res. 2008;19(1):9–...
Boronat López A, Balaguer Martínez J, Lamas Pelayo J, Carrillo García C, Peñarrocha DM. Resonance frequency analysis of dental implant stability during the healing period. Med Oral Patol Oral Cir Bucal. 2008;13(4):E244–7.
Gapski R, Wang HL, Mascarenhas P, Lang NP. Critical review of immediate implant loading. Clin Oral Implants Res. 2003;14(5):515–27.
Esposito M, Hirsch JM, Lekholm U, Th...
Schulte W, Lukas D. The Periotest method. Int Dent J. 1992;42(6):433–40.
Meredith N, Book K, Friberg B, Jemt T, Sennerby L. Resonance frequency measurements of implant stability in vivo. A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla. Clin Oral Implants Res. 1997;8(3):226–33.
Meredith N. Assessment of i...
Aparicio C, Rangert B, Sennerby L. Immediate/early loading of dental implants. A report from the Sociedad Española de Implantes World Congress consensus meeting in Barcelona, Spain 2002. Implant Dent Relat Res. 2003;5:57–60.
Cochran DL, Moeton D, Weber HP. Consensus statements and recommended clinical procedures regarding loading protocols for endosseous dental implants. Int J Oral Maxillofac ...
Cone beam CT
Computed tomography
Digital Imaging and Communications in Medicine
Implant stability quotient
Insertion torque
Multi-slice CT
The purpose of this study was to evaluate the relationship between the insertion torque value and the ISQ value at the implant treatment using the current rough-surfaced implant. As a result, no significant relationship was found between the insertion torque value and the ISQ value. Also, it was suggested that the ISQ value was considered to be an important indicator for observing the treatment st...
As accurate CT attenuation was not measured due to the lower spatial resolution of CBCT compared with MSCT, a CBCT was recognized as unsuitable for evaluating bone quality. However, several groups have recently reported the potential use of CBCT systems as an apparatus for estimating bone quality. Isoda et al. described a high correlation between voxel values obtained by CBCT and IT of the implant...
In this study, we could not find a significant relationship between insertion torque value and ISQ value. However, insertion torque value is an important indicator for predicting the progress of implant treatment, and ISQ value is considered to be an important indicator for observing the treatment state of the implant. Currently, the insertion torque value is used as the major decision index for t...
Increases or decreases of ISQ values are explained as follows: The inserted dental implant body is supported by mechanical interdigitating force after surgery, but this interdigitating force will be reduced time-dependently by the effects of osteoclasts activation at the initial stage of the bone remodeling process, then osseointegration will be completed by an increasing contact area between the ...
The insertion torque value in this study showed broader (10 to 50 N cm) than the previous publication (Table 2) [22, 32], and the cause of reasons for the difference are as follows: Primary stability may be affected by the bone quantity and bone quality in the treatment area, the micro- and macro-level design of the implant body, and the accuracy of the surgical technique [18, 25]. In this stud...
According to the previous literature, the obtaining osseointegration is integral to the intraosseous stability of the implant body during the healing period [24]; moreover, the importance of postoperative assessment of the intraosseous stability of the implant has also been reported [10]. Intraosseous stability of the implant body is evaluated immediately after the implant insertion and during the...
Average ISQ tended to increase during the healing period in all IT groups (Fig. 5). Average ISQ of the low IT group was 59.81 at 0 week, increasing significantly after ≥ 8 weeks (P
A total of 33 implant bodies (8 in the maxilla, 25 in the mandible) were inserted into the 27 participants (11 men, 16 women), with the average age of 54.6 ± 12.2 years (range, 32–78 years). The average IT value was 32.7 ± 9.2 N cm (32.5 ± 11.6 N cm in the maxilla, 32.8 ± 8.5 N cm in the mandible). The diameter of the implant body was 4.4 mm in 20 (60.6%) and 3.8 mm...
The CBCT was performed using a 3DX Multi-Image Micro CT FPD 8 system (J. MORITA MFG., Kyoto, Japan) (tube voltage, 80 kV; imaging area, 80 × 80 mm), and voxel values were measured with coDiagnostix™ 9.7 (dental wings, Montreal, Canada). The voxel values were calculated based on CT images for bone quality diagnosis. Voxel values were measured three times at 12 locations covering the mesial...
Implant treatment was performed in accordance with the procedure recommended by the manufacturer, without bone augmentation. A healing abutment was connected to the implant bodies after insertion (implant insertion in one stage method). A total of 17 dentists (treatment experience, 5–35 years; average, 11.5 years) performed all implant treatments in this study. All dentists who performed the i...
This prospective study was conducted jointly by Tokyo Dental College (Tokyo, Japan) and Fukuoka Dental College (Fukuoka, Japan) from January to December 2015. All study protocols were conducted in accordance with the Declaration of Helsinki [21] and were approved by the ethics committees of Tokyo Dental College (approval #416) and Fukuoka Dental College (approval #213).
Participants comprised pat...
The purpose of this study is to evaluate the relationship between IT and ISQ at implant treatment using the current rough surfaced implant. We evaluated the implant treatment sites with implant stability quotient (ISQ) values, IT values, and voxel values. We assumed that there is relevance between the insertion torque value and the ISQ value.
Dental implant treatments have improved in both convenience and predictability with refinements in implant bodies and treatment procedures as compared to about 50 years ago when clinical applications were started. Currently, an implant body surface is treated with “rough processing” by sandblasting and acid etching for the purposes of obtaining more reliable osseointegration and shortening tr...
A current implant body surface was treated with “rough processing” by sandblasting and acid etching for the purposes of obtaining more reliable osseointegration and shortening the treatment period. Various reports have examined the healing period with the use of these implant bodies, but a consensus opinion has not yet been obtained. The purpose of this study is to evaluate the relationship be...
A current implant body surface was treated with “rough processing” by sandblasting and acid etching for the purposes of obtaining more reliable osseointegration and shortening the treatment period. Various reports have examined the healing period with the use of these implant bodies, but a consensus opinion has not yet been obtained. The purpose of this study is to evaluate the relationship be...
Number of implants
Insertion torque value (N cm)
Implant stability quotient value
0 week
2 weeks
4 weeks
6 weeks
8 weeks
12 weeks
1
25
33.0
75.0
77.0
78.3
79.7
77.0
2
40
68.0
70.3
70.0
72.0
75.7
75.3
3
40
78.3
77.0
78.0
78.7
80.0
80.0
4
35
74.0
43.0
61.0
73.0
75.7
80.0
5
45
85.3
85.7
84.0
83.3
84.0
83.0
6...
Number of implants
Treatment area (FDI)
Size of implant (mm)
Length
Diameter
1
14
10
3.8
2
14
10
3.8
3
14
10
3.8
4
16
8
3.8
5
16
10
4.4
6
16
8
4.4
7
16
8
4.4
8
17
10
4.4
9
36
10
3.8
10
36
10
4.4
11
36
8
3.8
12
36
10
3.8
13
36
12
4.4
14
36
10
4.4
15
36
10
4.4
...
Figure 9. The comparison of two groups at average voxel values for each part. The comparison of voxel values by insertion torque. All specimens were classified into two groups by insertion torque
Figure 8. The relationship between average voxel value and insertion torque (averaged over the entire treatment area). The comparison of average voxel value among IT groups. Average voxel value was 384.0 ± 154.6 in the low IT group, 387.7 ± 147.7 in the medium IT group, and 619.2 ± 200.4 in the high IT group
Figure 7. The average voxel value between the maxilla and mandible. There was no difference between the maxilla (430.9 ± 211.6) and the mandible (475.6 ± 211.5) in the average voxel value. Also, no difference was found in each part
Figure 6. The relationship between ISQ and insertion torque. Percentage of specimens showing ISQ ≥ 73 compared with groups by week. In all groups, a period of rapidly increasing percentages was observed (8–12 weeks in the low IT group, 4–6 weeks in the medium and high IT groups). In the medium and high IT Group, a statistically significant difference was observed between ISQ ...
Figure 5. The comparison of ISQ values by the insertion torque. Time-lapse migration of ISQ values was compared with IT groups. Each IT group displayed similar migration. A significant difference in The ISQ was found in the low IT group after 8 weeks
Figure 4. The classification of the insertion torque. All specimens classified into three groups according to insertion torque. Criteria for the classification are shown in the figure and in the “Methods” section
Figure 3. The evaluation of the average ISQ. Time-lapse migration of average ISQ. Average ISQ of all specimens increased in a time-dependent manner (results indicated by a line). A significant difference was observed by 6 weeks after surgery
Figure 2. The measurement of the voxel values. A case of bone quality diagnosis before treatment. Width and height of the bone were measured to select the proper size of the implant body. The selected implant body was simulated on the bone images as a symbol, and then the voxel value was calculated as described in the "Method" section
Figure 1. Genesio® Plus implant with Aanchor surface. Scheme of the dental implant body for the Genesio® Plus implants with Aanchor surface used. a Overview picture of Genesio® Plus implants with Aanchor surface. b Image from scanning electron microscopy. Both pictures were provided by GC Corporation. To obtain osseointegration from an early stage, the dental implant body was treated...
According to the measurement of the average voxel values in this study, a significant difference was seen between the high IT group and the low/medium IT group, but no significant difference was found between the low and medium IT groups (Fig. 7). Specimens showing IT ≥ 40 N cm were thought to have a good bone quality, and voxel values at each part of the implant (neck, middle apex) were...
A bone quality of the treated area may affect primary stability as described above, preoperative analysis of bone quality is important for clarifying the primary stability of dental implants. This study analyzed bone quality using voxel values obtained using Digital Imaging and Communications in Medicine (DICOM) data from CBCT. According to the result of that analysis, it was...
A significant difference was observed between 0 and ≥ 8 weeks (Fig. 5). The ISQ did not change significantly during the experimental period in the medium or high IT groups, but the percentage of high ISQ (≥ 73) specimens was significantly higher at 4 to 6 weeks compared to other time periods in both groups (Figs. 5 and 6). The results in this study suggest that if the...
Other studies have suggested that ISQ immediately after implant insertion should be about 60, with ISQ subsequently decreasing over weeks 0–4 and increasing over weeks 4–8 after surgery. ISQ values 57–70 may indicate that intraosseous stability of the implant body is constant.
Increases or decreases of ISQ values are explained as follows: The inserted dental implant bo...
The insertion torque value in this study showed broader (10 to 50 N cm) than the previous publication (Table 2), and the cause of reasons for the difference are as follows: Primary stability may be affected by the bone quantity and bone quality in the treatment area, the micro- and macro-level design of the implant body, and the accuracy of the surgical technique. In this ...
Discussion
According to the previous literature, the obtaining osseointegration is integral to the intraosseous stability of the implant body during the healing period; moreover, the importance of postoperative assessment of the intraosseous stability of the implant has also been reported. Intraosseous stability of the implant body is evaluated immediately after the implant ...
Average ISQ tended to increase during the healing period in all IT groups (Fig. 5). Average ISQ of the low IT group was 59.81 at 0 week, increasing significantly after ≥ 8 weeks (P
Methods
Research design and study participants
This prospective study was conducted jointly by Tokyo Dental College (Tokyo, Japan) and Fukuoka Dental College (Fukuoka, Japan) from January to December 2015. All study protocols were conducted in accordance with the Declaration of Helsinki and were approved by the ethics committees of Tokyo Dental College (approval #416) and Fukuoka Dental Colle...
Results
Study overview
A total of 33 implant bodies (8 in the maxilla, 25 in the mandible) were inserted into the 27 participants (11 men, 16 women), with the average age of 54.6 ± 12.2 years (range, 32–78 years). The average IT value was 32.7 ± 9.2 N cm (32.5 ± 11.6 N cm in the maxilla, 32.8 ± 8.5 N cm in the mandible). The diameter of the implant body wa...
Methods
Research design and study participants
This prospective study was conducted jointly by Tokyo Dental College (Tokyo, Japan) and Fukuoka Dental College (Fukuoka, Japan) from January to December 2015. All study protocols were conducted in accordance with the Declaration of Helsinki and were approved by the ethics committees of Tokyo Dental College (approval #416) and Fuku...
Background
Dental implant treatments have improved in both convenience and predictability with refinements in implant bodies and treatment procedures as compared to about 50 years ago when clinical applications were started. Currently, an implant body surface is treated with “rough processing” by sandblasting and acid etching for the purposes of obtaining more reliable osseo...
Abstract
Background
A current implant body surface was treated with “rough processing” by sandblasting and acid etching for the purposes of obtaining more reliable osseointegration and shortening the treatment period. Various reports have examined the healing period with the use of these implant bodies, but a consensus opinion has not yet been obtained. The purpose of this study is to eval...